In-vitro Diagnostic

Hematology Analyzer

Features:

--3-part differentiation of WBC

--Double Chambers for WBC and RBC/PLT

--Throughout 60 samples per hour

--Intelligent reagent management function: Reagent information and residual display

--Strong data management system

--Multilingual for option

--LCD with high resolution

--Bar code scanner (optional)

--Reagent alarm system (optional)

 

Technical Specifications:

21Parameter + 3Histogram

WBC, LYM#, MID#, GRA#, LY%, MID%, GR%, RBC, HGB,HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT,

P-LCR, P-LCC and Histogram for WBC, RBC, PLT

Method:

Coulter Principle (Electrical impedance) Colorimetry hemoglobin determination

Performance

Parameter

Linearity range

Precision (CV%)

WBC(109/L)

0.0-100.0

2.0(at 7.0)

RBC(1012/L)

0.00-9.99

1.5(at 5.00)

HGB(g/L)

0-300

1.0 (at 140)

PLT(109/L)

0-1000

4.5(at 250)

 

 

Parameter

Linearity range

Precision (CV%)

WBC(109/L)

0.0-100.0

2.0(4.0-10.0)

RBC(1012/L)

0.00-9.99

1.5(3.50-5.50)

HGB(g/L)

0-300

1.0 (110-160)

PLT(109/L)

0-1000

4.5(100-300)

 

Throughout

60 samples per hour

Menu

Review, Test, Maintenance, QC, Setup, Shutdown, Prime reagents, System diagnosis, Aperture rinse, Soak and Rinse, Calibration, instrument config Display

8” color touch screen display

Sample volume Predilute:        20µL Whole blood:   10µL Aperture Diameter WBC    80µM

RBC/PLT            70µM

Carryover

WBC≤1.5%, RBC, HGB<1.0%, PLT≤2%

Input / Output

1 RS232, 1 keyboard, Mouse

4 USB, LAN

Printout

Built-in thermal printer Operating Environment Temperature:      10-35℃

Humidity:                  ≤90%

Power Requirement

AC 100-240V 50/60Hz

Dimension

Instrument:440mm(H) x 350mm(W) x 440mm(D) Package: 570mm(H) x 520mm(W) x 600mm(D) Net Weight

Net:17kg        Gross:21.8kg

Automatic Chemistry Analyser

Technical Specifications 

System Functions 

- Fully automatically, discrete, random STAT function 

- Online test items: at most 50 testing items 

- Throughput: 300 tests/hour 

- Analysis Method: end-point, kinetic, two-point, double 

- reagents,double-wavelength,multi-standard etc,open to various reagents 

- Data Processing: memory up to 2 million patients data 

Sample/Reagents Handling 

-Sample Position: 60 pcs ( Include standard, QC, STAT positions )

- Sample Probe Cleaning: automatic internal & external washing

- Sample Dilution: automatic/manual pre-dilution,with dilution ratio up to 1:100 

- Reagent Probe: automatic liquid level detection with collision protection function 

- Reagent Probe Cleaning: automatic internal & external washing 

Reaction System 

- Reaction Cuvette: 90 pcs made of special material 

- Optical Length of Cuvette: 6 mm

-Reaction Time : 8 -14 minutes

- Mixing System: independent mixing probe 

 

Optical System 

- Light Source: halogen-tungsten lamp

- Resolution : 0.001Abs 

- Spectrophotometry : rear spectrophotometry 

Dimension and Weight  

-Upper cabinet : 950x710x1180mm

Net Weight: 131 kgs

QC

- Various QC management function and random QC insert 

Calibration 

- Calibration : linear/nonlinear multi-points calibration 

- Re-test: retest the sample automatically when the result 

is out of the linearity range or the sample is not sufficient. 

 

Cleaning System 

- 8 step automatic washing, automatic cuvette dry function Water Consumption : 6L/hour

Head office (Corporate Office ): TRUTH BIOMEDICAL, C/o: AWFIS, 2A, 6TH Floor, Ecospace Business Park Premises, AA II, Rajarhat, Kolkata, West Bengal - 700156 , INDIA

whatsapp.png

+916289380385

Subscribe to Our Site

Copyright@2020 TRUTH BIOMEDICAL  - All Right Reserved. Copying any Details from this Website and Publish to any Other Website is a Legal Offence.