Hemodialysis Machine

Technology Data

Weight: 66kg

Input power: 1500w Standby battery: 30 minutes

Water inlet pressure: 0.15 MPa ~ 0.6 MPa

Water inlet temperature: 5℃ ~ 30℃

Working environment: Temperature 10℃ ~ 35℃ at relative humidity of less than 70%

Disinfection and Cleaning procedures

Chemical decalcification: Room temperature/85℃, flow: 300 ~ 800ml/min (adjustable)

Chemical    disinfection:   Room    temperature,    flow:    300    ~   800ml/min (adjustable)

Heat disinfection: 85℃, flow: 300 ~ 800ml/min (adjustable)

Intelligent disinfection setting from Monday to Sunday makes disinfection safer and more reliable.

Dialysate

Dialysate temperature: Preset range 33.0℃ ~ 39.0℃,±0.1℃; Dialysate flow: Less than 800ml/min (Adjustable)

Concentration of dialysate: 12.1 mS/cm ~ 16.0 mS/cm,±0.1 mS/cm; Dialysate ratio: Various proportions

UF(dehydration): 0 ml/h ~ 4000 ml/h Resolution: 1ml

Precision: ±1%

 

 

Heparin Pump

Syringe size: 20, 30, 50ml Flow range: 0 ml/h ~ 10 ml/h

First dose range: 1.0 ~ 20.0ml Resolution: 0.1ml

Precision: ±0.2 ml or ±5% of the reading

 

Extracorporeal circulation

Venous pressure: -180 mmHg  ~ +600 mmHg,  ±10 mmHg

Arterial pressure: -380 mmHg  ~ +400 mmHg,  ±10 mmHg

TMP pressure: -180 mmHg ~ +600 mmHg, ±20 mmHg

Blood pump, assistant pump flow range: 30 ml/min ~ 600 ml/min(Inner diameter Ф 8 mm)

Resolution: 1ml

Precision: ±10ml/min or 10% of the reading Blood leak detection: Photo chromicdetecting Bubble detection: Double protection

Head office (Corporate Office ): TRUTH BIOMEDICAL, C/o: AWFIS, 2A, 6TH Floor, Ecospace Business Park Premises, AA II, Rajarhat, Kolkata, West Bengal - 700156 , INDIA

whatsapp.png

+916289380385

Subscribe to Our Site

Copyright@2020 TRUTH BIOMEDICAL  - All Right Reserved. Copying any Details from this Website and Publish to any Other Website is a Legal Offence.